STUDI ANALISIS KETIDAKPASTIAN HASIL KALIBRASI TIMBANGAN DAN MISTAR TERHADAP KEBERTERIMAAN PENGUJIAN GRAMATUR KERTAS (STUDY ON ANALYSIS OF UNCERTAINTY OF CALIBRATED VALUE FOR BALANCE AND RULER FOR THE ACCEPTANCE OF PAPER GRAMMAGE TESTING)

Darmawan Darmawan(1*), Titik Istirohah(2)
(1) Balai Besar Pulp dan Kertas (bbpk)
(2) Universitas Negeri Malang
(*) Corresponding Author
DOI: http://dx.doi.org/10.25269/jsel.v6i02.99

Abstract

Study on analysis of uncertainty of calibrated value for balance and steel ruler for the acceptance of paper grammage testing that refers to SNI ISO 536 : 2010 (grammage testing method) and SNI 14-0440-2006 (paper and carton grammage) has been done. Testing instrument must be calibrated to ensure the testing value/measurement appropriate with expected specification refers to ISO 17025 : 2008. From calibration value gained conventional truth value and uncertainty of testing instrument. Uncertainty calculation refers to “Guide to the Expression of Uncertainty in Measurement”- ISO GUM, is a reference and measurement uncertainty evaluation method that have been received internationally. This reference is also one of reference documents in ISO 17025 : 2008 (competence requirements of examination and calibration laboratory). The evaluation test showed that the deviation value is less than 4%, so that it complies with the allowed tolerance which is also 4%. The result of testing evaluation and calibration showed that balance and ruler were used in Paper Testing Laboratory of Center for Pulp and Paper was appropriate with grammage testing prerequirement.




ABSTRAK


Studi analisis ketidakpastian hasil kalibrasi timbangan dan mistar baja terhadap keberterimaan uji gramatur kertas yang mengacu pada SNI ISO 536 Cara uji gramatur dan SNI 14-0440-2006 Gramatur kertas dan karton telah dilakukan. Untuk menjamin hasil uji/ pengukuran sesuai dengan spesifikasi yang diinginkan menurut ISO 17025 Syarat-syarat kompetensi Laboratorium Pengujian dan Laboratorium Kalibrasi maka alat uji/ukur harus dikalibrasi. Dari hasil kalibrasi diperoleh nilai kebenaran konvensional alat ukur beserta ketidakpastiannya. Perhitungan ketidakpastian mengacu pada ISO GUM “Guide to the Expression of Uncertainty in Measurement”, yang merupakan panduan dan metode evaluasi ketidakpastian pengukuran yang diterima secara internasional. Acuan ini juga sebagai salah satu dokumen acuan dalam ISO 17025 . Dari hasil evaluasi hasil uji diperoleh penyimpangan kurang dari 4%, sehingga memenuhi toleransi yang diizinkan yaitu 4 %. Kesimpulan dari studi ini menjelaskan bahwa timbangan dan mistar baja yang digunakan oleh Laboratorium Uji Kertas Balai Besar Pulp dan Kertas memenuhi syarat uji gramatur. Kata kunci: kalibrasi, ketidakpastian, gramatur, akurasi, toleransi

Kata kunci: kalibrasi, ketidakpastian, gramatur, akurasi, toleransi

Keywords

calibration; uncertainty; grammage; accuracy; tolerance

Full Text:

PDF

References

Benke, K. K., Lowell, K. E. and Hamilton, A. J. (2008) ‘Parameter uncertainty, sensitivity analysis and prediction error in a water-balance hydrological model’, Mathematical and Computer Modelling, 47(11–12), pp. 1134–1149.

Caciotta, M. (2008) ‘Informative calibration of the instrumentation’, Measurement: Journal of the International Measurement Confederation, 41(2), pp. 211–218.

Cox, M. G. and Harris, P. M. (2006) ‘Measurement uncertainty and traceability’, Measurement Science and Technology, 17(3), pp. 533–540.

ISO (2008) ‘Guide 98-3. Uncertainty of measurement - Part 3: Guide to the expression of uncertainty in measurement (GUM:1995)’, Guide 98-3, p. 130.

JCGM 100 (2008) ‘JCGM 100:2008 - Evaluation of measurement data - Guide to the expression of uncertainty in measurement’, International Organization for Standardization Geneva ISBN, 50(September), p. 134.

Kammeyer, M. E. and Rueger, M. L. (2008) ‘Estimation of the Uncertainty in Internal Balance Calibration Through Comprehensive Error Propagation’, 26th AIAA Aerodynamic Measurement Technology and Ground Testing Conference, (June), pp. 1–16.

Les Kirkup, B. F. (2006) ‘An Introduction to uncertainty in measurement using the GUM (Guide to the expression of uncertainty in measurement)’, Accreditation and Quality Assurance, 12(1), p. 249.

Malengo, A. (2014) ‘Buoyancy effects and correlations in calibration and use of electronic balances’, Metrologia, 51(5).

Mayr, E., Hagg, W., Mayer, C. and Braun, L. (2013) ‘Calibrating a spatially distributed conceptual hydrological model using runoff, Annual mass balance and winter mass balance’, Journal of Hydrology, 478, pp. 40–49.

Meyer, V. R. (2007) ‘Measurement uncertainty’, Journal of Chromatography A, pp. 15–24.

Mishra, S. (2009) ‘Uncertainty and sensitivity analysis techniques for hydrologic modeling’, Journal of Hydroinformatics, 11, p. 282.

United Kingdom Accreditation Service (2007) ‘The Expression of Uncertainty and Confidence in Measurement’, Measurement, 44(January), p. 28. Available at: http://www.ukas.com/library/Technical-Information/Pubs-Technical-Articles/Pubs-List/M3003.pdf.


Article Metrics


Abstract view : 1172 times
PDF view : 988 times

Refbacks

  • There are currently no refbacks.


Copyright (c) 2016 JURNAL SELULOSA
Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License.